Total Chlorine and sulfur analyzer TCL 100
Analysis method: Oxidative decomposition, Coulometric titration
Combustion: Silica tube combustion
Combustion furnace temperature: 1100 °C at the maximum
Samples to analyze: Liquid and gaseous samples
Detection method: Oxidation-
Detecting electrode: Silver electrode for Chlorine and Alloy electrode for sulfur
Titration control: Automatic electrolytic current control
Amount of samples to measure: Liquid: 5 to 200 µl*1 Gas: 5 to 10 ml*2
Measurement range: 0.1 to 50 µg (Cl) 0.5 to 1,000 ppm
Measurement accuracy: Standard samples: CV of 5% or below for 1 µg of chlorine
When 2,4-
For sulfur: 0.02 to 20 µg (S) 0.1 to 1.000 ppm (Samples burnable below 1,000 °C with no difficulty)
Measurement time: Approx. 10 min./sample (Content of 2 ug at measurement)
Interface: 1 channel for PC input/output (RS232C)
Instrument control and Data processing:
Via PC software, application data bank, export functions, QC Check functions
Gas used:
Argon: Purity greater than 99.98%, 300±100 kPa
Oxygen: Purity greater than 99.7%, 300±100 kPa
Power supply:
Main unit: AC 100/115/230/240 V, 50/60 Hz, 1300 VA
Detector: AC 100/115/230/240 V, 50/60 Hz, 40 VA
Dimensions and weight:
Main unit: Approx. 500 (W) x 400 (D) x 500 (H) mm, approx. 35 kg
Detector: Approx. 220 (W) x 370 (D) x 415 (H) mm, approx. 8 kg