YL9101 Vacuum Degasser
Number of channel: 4 CHs
Maximum flow rate: 10ml/min per channel
> 0-2.0ml/min per channel for 70% Gas Removed from Methanol
Internal volume per channel: 925ul per channel
Materials in contact with solvent: Teflon AF, PEEK and Glass-filled PTFE
Safety & maintenance: Error detection
Dimensions: 385 X 80 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, automatic voltage selection; 50/60Hz, ±5%, 20W


YL9110 Quaternary Pump
Operating principle: Parallel dual-plunger pump, Low-pressure gradient
Control mode: Constant flow, Constant pressure
Compressibility compensation: Automatic
Flow range: 0.001-10ml/min
Flow rate accuracy : ≤1% at 1ml/min
Flow rate precision: 0.1% RSD at 1ml/min
Maximum pressure: 6000 psi > Operating range: 0-6000 psi up to 5ml/min > Operating range : 0-3000 psi up to 10ml/min
Pressure pulsation : ≤1% at 1ml/min
Number of eluent lines: 4
Gradient formation: 4-channel mixing valve
Composition Precision: <0.1%
Composition Accuracy: <0.5%
Auto prime/purge
Communications: LAN
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, automatic voltage selection, 50/60Hz, ±5% , 70W
YL9111 Binary Pump
Operating principle: X 2 Parallel dual-plunger pump, High-pressure gradient
Flow range: 0.001-10ml/min
Number of eluent lines: 2
Communications: LAN
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, 50/60Hz, ±5%, 140W
.YL9112 Isocratic Pump
Operating principle: Parallel dual-plunger pumpCompressibility compensation: Automatic
Flow range: 0.001-10ml/min
Flow rate accuracy : ≤1% at 1ml/min
Flow rate precision: 0.1% RSD at 1ml/min
Maximum pressure: 6000 psi
> Operating range: 0-6000 psi up to 5ml/min
> Operating range: 0-3000 psi up to 10ml/min
Pressure pulsation : ≤1% at 1ml/min
Composition Precision: <0.1%
Composition Accuracy: <0.5%
Auto prime/purge
Communications: LAN
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, 50/60Hz, ±5%, 70W
YL9120 UV/Vis Detector
Wavelength Range: 190-900 nm
Data collection rate: up to 50Hz
Deuterium arc lamp & tungsten lamp
2-Order filter: Automatic filter switching
Bandwidth: 5.5 nm
Wavelength Accuracy : ±1 nm
Wavelength Precision : ±0.1nm
Linearity: >99.5% at 2.5 AU (acetone, 254nm)
Noise level : <±0.5 X 10^(-5) Abs/min , 254nm, dry cell
Drift : <1 X 10^(-4) Abs/hour
Dispersion element: Concave grating
Warm-up period: 1 hour
Flow Cell Design: Cone type
Path Length: 10 mm (Analytical cell)
Cell Volume: 10 uL (Analytical cell)
Pressure limit: 1500psi (Analytical cell)
Wetted materials: 316 stainless steel, Quartz, Teflon
Analog Outputs: Two, software selectable: Absorbance Energy, Sample Energy, Reference Energy Communications: LAN
Power-up Diagnostics: Optics and electronic diagnostic routine
Power-up Wavelength Verification: Automatic on power up via internal holmium filter and D2 lamp
5-Point Wavelength Calibration: On demand via internal holmium filter and D2 lamp
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, 50/60Hz, ±5%, 100W
YL9160 PDA Detector
PDA A/D Resolution: 16bit
Slit Bandwidth: 1.7nm Pixel Resolution: 0.9nm Channel: 1024
Wavelength: 190~950
Shield Optics
Analytical Cell
- Path-length: 10mm
- Pressure : >1500psi
- Volume : 13ul
Dimension: W: 385mm H: 160mm D: 565mm
Weight: 15Kg
Noise Level: +/- 0.00002 AU (Empty Cell, 1sec Rise Time, 254nm)
Drift : 0.0002AU/hr (Baseline Correction), 0.001AU/hr (Room Temp)
Wavelength Accuracy: <1nm (HY-1 Holmium Oxide Filter)
Sampling Rate: Max. 50Hz 1~16 nm Spectral Bandwidth, Wide Spectral Range, Alternative Spectral Function
GLP Compliance:
- Photometric Accuracy, Linearity, Noise Level, Drift - System Check,
Analog Output: 2 channel (Replaceable Dual Wavelength Detector)
Valve Output: 2ch, Programmable Sampling (Fraction Collector Function Available )
Trigger Input: 2ch
Communication: TCP/IP, RS232C
Simultaneous Data Channel: 8 or Full Scan
Filtering: Bessel, RC, Aver. 0.01~10Hz
YL9150 Autosampler (Alias)
Sound pressure level: LeAq < 70 dB
Working temperature: 10 - 40°C
Humidity: 20 - 80% RH
Safety and EMC compatibility: According to EC-directives; CSA (UL) approved
Power requirements: 95 - 240 Volt AC ± 10%; 50/60 Hz; 200VA
Sampling:
Sample capacity: 2 Micro Well Plates according to SBS standards; 96-well high/low
Loop volume: 1 - 5000 µL programmable, 10 mL loop optional
Dispenser syringe: 500 µL standard or 2500 µL for Prep option
Vial detection: Missing vial/well plate detection by sensor
Headspace pressure: Built-in compressor, but only for vials with septa
Switching time injection valve: Electrically < 100 msec
Piercing precision needle: ± 0.6mm
Wash solvent: External wash solvent bottle (bracket optional)
Wetted parts in flow path: SS316, PTFE (teflon), ETFE (tefzel), VESPEL, PCTFE (Kel-F). Optional: PEEK
(10) Injection cycle time: < 60 sec. in all injection modes for 1 injection ≤ 100 µL including 300 µL wash / < 20 sec in partial loopfill mode for 1 injection of 10µL including 500 µL wash
Performance
Easy to use
Fast installation
What is your choice tomorrow ?
Contact us NOW for a quote
Our new ELSD is an essentially universal detector, primarily used in High Performance Liquid Chromatography (HPLC). Young Lin ELSD is an ideal substitute,or supplement to, traditional HPLC detectors for liquid chromatography concentration detection.
In comparison to other ELSD detectors, Young Lin ELSD has the ability to vary the
split ratio smoothly over a wide range with a new patented (Patent No. US 7,290,723
B1) Thermo-Split Technology. This precise vapor phase control combines a gentle bend
with a temperature controlled spray chamber.
This means that with one low cost detector
you can handle high volume as well as difficult to evaporate mobile phases optimally,
efficiently and easily, all without sacrificing sensitivity.
Features
Sensitive Detection for a Wide Variety of Chemical Classes and Structures
Patented Thermo-Split™ Technology for Precise Vapor Phase Control
(Patent No. US 7,290,723 B1)
A single multi-flow, nebulizer for use with micro-bore to semi-preparative flow rates.
Very low detector volume resulting in the smallest peak within 3 sec
Fast Start-up / Easy to Use
No Consumable Parts
Low Cost of Operation
There are 2 types of ELSD as below.
YL9180 ELSD
Display 2 Line x 20 Character per line VFD
User Interface Two multi-function buttons Four multi-function buttons
Evaporative Zone Temp. 60°C Ambient to 120°C
Thermo-Split™ Chamber Temp. 30°C to 60°C
Liquid Flow Rate 0.2mL/min to 3mL/min
Wetted Materials Stainless steel, glass, anodized aluminum, Teflon™
Light Source 670 nm Laser Diode, <5mW
Detector Hermetically sealed photo-diode/operational amplifier
Interface RS232, Contact Closure
Typical Laboratory General Laboratory Advanced Research
Sensitivity More Sensitive
YL9181 ELSD
Display 2 Line x 20 Character per line VFD
User Interface Four multi-function buttons
Evaporative Zone Temp. 60°C Ambient to 120°C
Thermo-Split™ Chamber Temp. 10°C to 70°C
Liquid Flow Rate 0.2mL/min to 5mL/min
Wetted Materials Stainless steel, glass, anodized aluminum, Teflon™
Light Source 670 nm Laser Diode, <5mW
Detector Hermetically sealed photo-diode/operational amplifier
Interface RS232, Contact Closure
Typical Laboratory General Laboratory Advanced Research
Sensitivity Highly Sensitive