TS-100 Total Sulfur analyzer with Total nitrogen option

for Solid, liquid and gaseous samples

Using high performance UV fluorescence method for measurement of sulfur content in solid, liquid and gaseous samples down to the ppb level, this instrument is the best choice for his sensitivity and performances.

Fully compatible with the ASTM D5453, 3246JIS K 2609, ASTM - D 4629, D5176,D5762 and D6069

TS-100 V automated02

Measurement sample             Solid, Liquid and gaseous sample (Use an option )

Measurement substance        Sulfur compound and Nitrogen compound ( ND-100 option )

Analysis method                      Oxidative pyrolysis , ultraviolet fluorescence method  and  chemioluminescence  for ND  100 option Sample injection method        Direct injection by micro syringe

Optional Model :                      CRI-100H ; Gas injector GI 100; LPG injector GI 300 ; Liquid  autosampler ASC-150L and Solid                                                                                                                            sampler ASC 120 S)

Furnace temperature              1000 - 1100 degre C (It can be set separately at sample injection part  and combustion part.)

 

Measurement range        

  

Measurement time                   3 to 10 minutes every measurement

Data processing:                      by PC and Software; application database and excel transfer

Detector output signal              DC 0 to 1V

Balance input                            RS232C

Gas                                            Oxygen gas (Purity : more than 99.7 %, 0.3±0.1MPa)

                                                   Argon gas (Purity : more than 99.98%, 0.3±0.1Mpa)

 

Standard use amount is as follows:

 

Oxygen gas :                            400ml/min

Argon gas (for analysis) :        500ml/min

Argon gas (for purge) :            650ml/min

Supply gas pressure fluctuation : under ±2% in one day

Power TS-100V :                    AC 100/115/230/240V, 50/60 Hz, 1400VA
SD-100 :                                   AC 100/115/230/240V, 50/60 Hz, 52VA

Dimension and weight:


TS-100 main unit :                   Approx. 500(W)×360(D)× 430(H) mm,   Approx. 33 kg
Sulfure detector:                      Approx. 260(W)×360(D)× 430(H) mm,   Approx. 19 kg
Automatic Boat Controller :    Approx. 440(W)×250(D)× 180(H) mm,   Approx. 11 kg

ND option:                                Approx. 260(W)×360(D)× 430(H) mm,   Approx. 17 kg

Sensitivity: 

Solid

Liquid

Gas (ml/m3)

Sulfur

0.05 to 10000  ppm

0.05 to 10000 ppm

0.05 to 10000 ppm

Nitrogen

0.5 to 5000 ppm

0.5 to 5000 ppm

0.5 to 5000 ppm

Sample quantity

1- 30 mg

5 - 50µl

1 to 10ml

 