TN-100 Total nitrogen analyzer

for liquid and gaseous samples

TN-100 automated

The TN-100 is provided with all the accessories required for measurement of non-aqueous and aqueous samples The use of the model ASC-150L liquid auto sampler allows completely automated measurement of up to 50 samples.A reduced pressure system is used that is essential for non-aqueous samples.

Five types of measuring ranges make it possible to accommodate a wide range of sample types from low concentrations to high concentrations. Safe design featuring continuous monitoring of supplied gas flow rate, reaction tube temperature and other parameters along with activation of an interlock function during system abnormalities.

Applicable samples Aqueous, non-aqueous and gaseous samples

Analysis method Oxidative pyrolysis, chemiluminescence method (reduced pressure)

Catalyst:  Platinum oxidation catalyst

Furnace temperature : Max 999°C

Measuring ranges:   0.01 - 5,000 ppm

Sample size: 200 ul ore less

Measurement time: Within 10 min / measurement

 

Reproducibility ( when using standard sample) :

     

Analog output : DC 0-1 V (full scale)

Gas requirements: Argon (purity: 99.98% or more, 290-490 kPa) 

Oxygen (purity: 99.7% or more, 290-490 kPa )

Power requirements: TN-100 : AC 100/115/230/240 V, 50/60 Hz, 1,000 VA

Dimensions and weight: TN-100 : Approx. 560(W) x 400(D) x 520(H) mm, approx. 53 kg

Aqueous samples:

   Concentration

     Sample size 

    RSD (%) n=5

  Using ammonium sulfate

 

  0.1-1.0  

100          

  < 8

 

 

  1.0 - 5,000

50

 < 3

 

Non-aqueous samples 

 

 

 

Using pyridine  toluene

 

0.1-1.0 

100 

  < 8

 

 

1.0 - 5,000

20 -100

  < 3

 