YL9101 S Vacuum Degasser
Number of channel: 4 CHs
Maximum flow rate: 10ml/min per channel
> 0-2.0ml/min per channel for 70% Gas Removed from Methanol
Internal volume per channel: 925ul per channel
Materials in contact with solvent: Teflon AF, PEEK and Glass-filled PTFE
Safety & maintenance: Error detection
Dimensions: 385 X 80 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, automatic voltage selection; 50/60Hz, ±5%, 20W
YL9100 S HPLC Semi Prep HPLC


YL9110 S Quaternary Pump
Operating principle: Parallel dual-plunger pump, Low-pressure gradient
Control mode: Constant flow, Constant pressure
Compressibility compensation: Automatic
Flow range: 0.01-50ml/min
Flow rate accuracy : ≤1% at 5ml/min
Flow rate precision: 0.1% RSD at 5ml/min
Maximum pressure: Operating range: 0-3500 psi up to 35ml/min > Operating range : 0-2500 psi up to 35 - 50ml/min
Pressure pulsation : ≤1% at 25ml/min
Number of eluent lines: 4
Gradient formation: 4-channel mixing valve
Composition Precision: <0.2% at 5 ml/ min
Composition Accuracy: <0.3% at 5 ml/min
Semi Auto prime/purge
Communications: LAN
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, automatic voltage selection, 50/60Hz, ±5% , 70W
YL9111 S Binary Pump
Operating principle: X 2 Parallel dual-plunger pump, High-pressure gradient
Flow range: 0.01-50ml/min
Number of eluent lines: 2
Communications: LAN
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, 50/60Hz, ±5%, 140W
.YL9112 S Isocratic Pump
Operating principle: Parallel dual-plunger pump
Compressibility compensation: Automatic
Flow range: 0.01-50ml/min
Flow rate accuracy : ≤1% at 5ml/min
Flow rate precision: 0.1% RSD at 5ml/min
Maximum pressure: 6000 psi
> Operating range: 0-3500 psi up to 35ml/min
> Operating range: 0-2500 psi from 35 ml/min up to 50ml/min
Pressure pulsation : ≤1% at 1ml/min
Composition Precision: <0.1%
Composition Accuracy: <0.5%
Auto prime/purge
Communications: LAN
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, 50/60Hz, ±5%, 70W
YL9120 S UV/Vis Detector
Wavelength Range: 190-900 nm
Data collection rate: up to 50Hz
Deuterium arc lamp & tungsten lamp
2-Order filter: Automatic filter switching
Bandwidth: 5.5 nm
Wavelength Accuracy : ±1 nm
Wavelength Precision : ±0.1nm
Linearity: >99.5% at 1.0 AU (acetone, 254nm)
Noise level : <±0.5 X 10^(-5) Abs/min , 254nm, dry cell
Drift : <3 X 10^(-4) Abs/hour
Dispersion element: Blazed Concave holographic grating
Warm-up period: 1 hour
Flow Cell Design: Cone type
Path Length: 3 mm (Semi prep cell)
Cell Volume: 5 uL (Sedmi prep cell)
Pressure limit: 1500psi (Semo prep cell)
Wetted materials: 316 stainless steel, Quartz, Teflon
Analog Outputs: Two, software selectable: Absorbance Energy, Sample Energy, Reference Energy Communications: LAN
Power-up Diagnostics: Optics and electronic diagnostic routine
Power-up Wavelength Verification: Automatic on power up via internal holmium filter and D2 lamp
5-Point Wavelength Calibration: On demand via internal holmium filter and D2 lamp
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, 50/60Hz, ±5%, 100W
YL9160 S PDA Detector
PDA A/D Resolution: 16bit
Slit Bandwidth: 1.7nm Pixel Resolution: 0.9nm Channel: 1024
Wavelength: 190~950
Shield Optics
Semi Prep Cell
- Path-length: 3 mm
- Pressure : <1500psi
- Volume : 5ul
Dimension: W: 385mm H: 160mm D: 565mm
Weight: 14Kg
Noise Level: +/- 0.00002 AU (Empty Cell, 1sec Rise Time, 254nm)
Drift : 0.0002AU/hr (Baseline Correction), 0.001AU/hr (Room Temp)
Wavelength Accuracy: <1nm (HY-1 Holmium Oxide Filter)
Sampling Rate: Max. 50Hz 1~16 nm Spectral Bandwidth, Wide Spectral Range, Alternative Spectral Function
GLP Compliance:
- Photometric Accuracy, Linearity, Noise Level, Drift, System Check,
Analog Output: 2 channel (Replaceable Dual Wavelength Detector)
Valve Output: 2ch, Programmable Sampling (Fraction Collector Function Available )
Trigger Input: 2ch
Communication: TCP/IP, RS232C
Simultaneous Data Channel: 8 or Full Scan
Filtering: Bessel, RC, Aver. 0.01~10Hz
YL9150 Autosampler (Alias)
Sound pressure level: LeAq < 70 dB
Working temperature: 10 - 40°C
Humidity: 20 - 80% RH
Safety and EMC compatibility: According to EC-directives; CSA (UL) approved
Power requirements: 95 - 240 Volt AC ± 10%; 50/60 Hz; 200VA
Sampling:
Sample capacity: 2 Micro Well Plates according to SBS standards; 96-well high/low
Loop volume: 1 - 5000 µL programmable, 10 mL loop optional
Dispenser syringe: 500 µL standard or 2500 µL for Prep option
Vial detection: Missing vial/well plate detection by sensor
Headspace pressure: Built-in compressor, but only for vials with septa
Switching time injection valve: Electrically < 100 msec
Piercing precision needle: ± 0.6mm
Wash solvent: External wash solvent bottle (bracket optional)
Wetted parts in flow path: SS316, PTFE (teflon), ETFE (tefzel), VESPEL, PCTFE (Kel-F). Optional: PEEK
(10) Injection cycle time: < 60 sec. in all injection modes for 1 injection ≤ 100 µL including 300 µL wash / < 20 sec in partial loopfill mode for 1 injection of 10µL including 500 µL wash

YL-Clarity Data System
Direct control of YL9100 HPLC.
Measuring: Simultaneous data acquisition from up to four independent chromatographs, each chromatograph can acquire data from up to 12 detectors.
Integration: There is extensive possibility to modify chromatograms. The chromatogram can be changed by entering global parameters or interactively, through direct graphic modification of the baseline.
Overlay: Simultaneously displays a virtually unlimited number of chromatograms and their mathematical modification; for example, mutual deductions or derivations of any order.
Calibration: Internal and external standard calculation methods, calibration of groups of peaks and reference peaks method for better identification.
Automated measuring support: Sequence tables for any set of samples with or without an autosampler.

Post-run: Automatically displays, prints, exports and starts other programs after the completion of a measurement.
Summary result tables: Displays and prints selected results from all simultaneously displayed chromatograms.
User settings: User selects parameters for peak display and the specification for axes, including color from an extensive array of color settings. Text labels and lines, either as part of the area or anchored to a chromatogram, may also be inserted.
Export: Optional exportation of all results, with or without the chromatogram, in various formats, into a file or clipboard.
Import: Imports chromatograms or mathematical curves, which have been saved in text or AIA formats, from other programs.
Method and calibration history: Each chromatogram can easily be displayed under the same conditions as when it was printed, exported or saved.
Column performance: Calculations of peaks in terms of symmetry, efficiency, resolution; all by several methods (tangent, moments, etc.).
Batch: Automatically batch processes, displays, exports or prints any number of chromatograms.

User calculations: Users can define custom calculations in the Result and Summary tables. Using the integrated editor you can create your own columns from original columns and individual mathematical functions.
User accounts: Sets up access rights and passwords (including their parameters e.g., minimum length, validity, etc.). Each user can define his or her own station appearance.
Audit trail: Records selected events and operations into a special file. Records selected operations directly into a chromatogram.
Electronic signature: Each chromatogram can be signed electronically. Signature selection is based on the username or the signature certificate.
Networked Solution: Clarity chromatography station files can be accessed from networked computers using Clarity software. This enables the offline evaluation of chromatograms, development of new methods and printing reports.
21 CFR Part 11: Clarity Satisfies with the requirements of the 21 CFR Part 11 directive of the FDA.
Operating Windows: Microsoft Window 2000, XP and Vista (32 bit)
Searching for Semi Prep ?
Ask fo a quote NOW
YL9170 S Refractive Index Detector
Measurement methode : deflection
RI Range: 1.00 to 1.75 RIU
Cell:
- Flow rate : 1-50 ml/min
- Pressure : 6 kg/cm2 / 84 psi
- Volume : 7 ul
Dimension: W: 385mm H: 160mm D: 565mm
Weight: 17Kg
Noise Level: <10 x 10 exp -8 RIU )
Drift with 1 ml Water/ min: 1mV/h
Linearity range : 0-20000uRID
Temperature control : 35 degree to 55 degree
Analog Output: + - 1 V
Communication: RS232C
Digital input: purge, Auto Zero, Start marker