YL9110 Quaternary Pump
Operating principle: Parallel dual-plunger pump, Low-pressure gradient
Control mode: Constant flow, Constant pressure
Compressibility compensation: Automatic
Flow range: 0.001-10ml/min
Flow rate accuracy : ≤1% at 1ml/min
Flow rate precision: 0.1% RSD at 1ml/min
Maximum pressure: 6000 psi > Operating range: 0-6000 psi up to 5ml/min > Operating range : 0-3000 psi up to 10ml/min
Pressure pulsation : ≤1% at 1ml/min
Number of eluent lines: 4
Gradient formation: 4-channel mixing valve
Composition Precision: <0.1%
Composition Accuracy: <0.5%
Auto prime/purge
Communications: LAN
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, automatic voltage selection, 50/60Hz, ±5% , 70W
YL9120 UV/Vis Detector
Wavelength Range: 190-900 nm
Data collection rate: up to 50Hz
Deuterium arc lamp & tungsten lamp
2-Order filter: Automatic filter switching
Bandwidth: 5.5 nm
Wavelength Accuracy : ±1 nm
Wavelength Precision : ±0.1nm
Linearity: >99.5% at 2.5 AU (acetone, 254nm)
Noise level : <±0.5 X 10^(-5) Abs/min , 254nm, dry cell
Drift : <1 X 10^(-4) Abs/hour
Dispersion element: Concave grating
Warm-up period: 1 hour
Flow Cell Design: Cone type
Path Length: 10 mm (Analytical cell)
Cell Volume: 10 uL (Analytical cell)
Pressure limit: 1500psi (Analytical cell)
Wetted materials: 316 stainless steel, Quartz, Teflon
Analog Outputs: Two, software selectable: Absorbance Energy, Sample Energy, Reference Energy Communications: LAN
Power-up Diagnostics: Optics and electronic diagnostic routine
Power-up Wavelength Verification: Automatic on power up via internal holmium filter and D2 lamp
5-Point Wavelength Calibration: On demand via internal holmium filter and D2 lamp
Safety & maintenance: Leak detection, Diagnostics, Error detection
Dimensions: 385 X 160 X 565mm (width X height X depth)
Line Voltage : 100-240VAC, ±10%, 50/60Hz, ±5%, 100W
YL9150 Autosampler (Alias)
Sound pressure level: LeAq < 70 dB
Working temperature: 10 - 40°C
Humidity: 20 - 80% RH
Safety and EMC compatibility: According to EC-directives; CSA (UL) approved
Power requirements: 95 - 240 Volt AC ± 10%; 50/60 Hz; 200VA
Sampling:
Sample capacity: 2 Micro Well Plates according to SBS standards; 96-well high/low
Loop volume: 1 - 5000 µL programmable, 10 mL loop optional
Dispenser syringe: 500 µL standard or 2500 µL for Prep option
Vial detection: Missing vial/well plate detection by sensor
Headspace pressure: Built-in compressor, but only for vials with septa
Switching time injection valve: Electrically < 100 msec
Piercing precision needle: ± 0.6mm
Wash solvent: External wash solvent bottle (bracket optional)
Wetted parts in flow path: SS316, PTFE (teflon), ETFE (tefzel), VESPEL, PCTFE (Kel-F). Optional: PEEK
(10) Injection cycle time: < 60 sec. in all injection modes for 1 injection ≤ 100 µL including 300 µL wash / < 20 sec in partial loopfill mode for 1 injection of 10µL including 500 µL wash
Simultaneous sample injection in up to 4 HPLC columns and parallel data treatment of four UV/VID detector signal

Contact us NOW for a quote