AQF-100  Automated sample mineralisation

for solid and liquid samples  

wpffabd1f7.gif
wpffabd1f7.gif

Automation for halogen and sulfur analysis in complex samples using an IC system

 

The AQF-100 is a completely  unique instrument for qualitative and quantitative measurement of all halogens and sulfur  in practically  any kind of samples.

This instrument can replace all the mineralisation systems (Oxygen  bomb, microwave high pressure mineralisation) and , in conjunction with a Ion Chromatography system  allow total automation in the analysis of complex samples, industrial  residue, solvent mixes, soil samples, resin, wood  etc...  Save time , increase performance, increase throughput, cut your analysis cost with this unique instrument.

 

Analysis method:                              Oxidative decomposition, Ion Chromatography (not provided)

Combustion:                                      Pyrolysis tube silica tube combustion

Combustion furnace temperature:  1100 °C  maximum

Samples to analyze:                         Liquid, solid and gaseous samples(with special sampling device)

Enrichment  method:                        Using trapping solution , multi enrichment possible

Amount of samples to measure:     Solid: up to 1-100 mg;  Liquid: 1 to 100 µl  ; Gas: 1 to 10 ml

Measurement range:                        depending on IC system, compatible with Metrohm IC and  Dionex IC Measurement accuracy check:       Using internal standard                          

Measurement time:                          Approx. 10-15 min./sample depending on IC method, including sample injection and                                                                             mineralisation

Interface:                                           1 channel for PC input/output (RS232C)

Instrument control and Data processing: Via PC software, application data bank, export functions, QC Check  functions

 

Gas used:

Argon: Purity greater than 99.98%, 300±100 kPa

Oxygen: Purity greater than 99.7%, 300±100 kPa

Power supply:

Main unit complete :                                            AC  100/115/230/240 V, 50/60 Hz, 1300 VA

 

 

Dimensions and weight:

Main unit:                                            Approx. 500 (W) x 440 (D) x 500 (H) mm,  24 kg

Absorption unit GA-100                    Approx. 250 (W) x 425 (D) x 560 (H) mm,  20 kg

Injection unit ABC                              Approx. 440 (W) x 250 (D) x 180 (H) mm,  7 kg

WS-100 unit:                                      Approx. 360 (W) x 145 (D) x 140 (H) mm,  4 kg












wp1e6c5cb9.png
wp9f49585d.png
wp92abdd90.gif
wp77c6cf66.png

Chlorine

Bromine

Iodine

Fluorine

Sulfur