Analysis method: 

 

Halogens :

Oxidative decomposition, High sensitive Coulometric  titration, Automatic electrolytic current control

Nitrogen:  

Chemioluminescence

Sulfur:  

UV Fluorescence

Combustion:  

 

Halogen

Silica tube combustion

Nitrogen

Silica tube combustion and catalyst

Sulfur

Silica tube combustion

Temperature

1100 °C at the maximum

Samples to analyze:

Solid, Liquid and gaseous samples

Measurement range

 

UV fluorescence Sulfur

Solid: 0.05 - 10000ug/g   Liquid: 0.05 - 5000 ug/ml

CLD -Nitrogen

Solid: 0.5 - 5000ug/g    Liquid: 0.2 - 5000 ug/ml

Coulometry Chlorine

0.01 - 500 ug ( 0,1 - 5000 ug/ml )

Coulometry  Sulfur

0.05 - 50 ug (0.5 - 500 ug/ml)

Sample size

 

Non aqueous sample

50 ul (or up to 100ul with option)

Solid

30 mg ( or up to 150 mg with option)

Measuring time

 

UVFL / CLD

3 - 10 min  simultaneous Sulfur and Nitrogen

Coulometry

< 10 min.

Gas

Argon and Oxygen

Software

 

Complete set

For Microsoft Windows 7 prof. 32 bit, including calibration  test and QC functions, reusable calibration curves, recalculation and GLP functions.

The complete solution for all elementary analysis 
used in environmental analysis and petrochemical industry,
NEW